  * Use images over a reasonable range of defocuses
  * Don't forget to pick a good amplitude contrast (10-20% works best for cryo data)
  * Use images over a reasonable range of defocuses
  * Don't forget to pick a good amplitude contrast (10-20% works best for cryo data)
Line 12: Line 12:
  * use the same %AC you used above
   * guess at a B-factor based on the fitting above. For a high-end 300 keV instrument this may be as low as 50-100. For a typical 200 FEG scope it will be in the 150-200 range,
  * use the same %AC you used above
  * guess at a B-factor based on the fitting above. For a high-end 300 keV instrument this may be as low as 50-100. For a typical 200 FEG scope it will be in the 150-200 range,
Line 15: Line 15:
  * run this program once for each stack of particles (one stack per frame). The particles should be unmasked and unfiltered !
  * fitctf <particles from one frame> sf=<sf file from above> apix=<A/pix> V=<voltage> Cs=<spher aber> acfix=<amp contrast, 0.1 is 10%> Bhint=<guess at EMAN1 style B-factor>
  * run this program once for each stack of particles (one stack per frame). The particles should be unmasked and unfiltered !
  * fitctf <particles from one frame> sf=<sf file from above> apix=<A/pix> V=<voltage> Cs=<spher aber> acfix=<amp contrast, 0.1 is 10%> Bhint=<guess at EMAN1 style B-factor>
Line 18: Line 18:
  * Run 'ctfit *.tnf' This will open the results of the automatic fitting.
  * If you need to correct any of the fit values (you very well may):
     * select the image in the upper right list in the control panel
  
* make necessary adjustments
     * press the 'set' button below the lower right list in the control panel, this will save your parameters to disk
  * Do NOT use the 'phase flip' menu item when working this way, since you loaded '.tnf' files instead of the original particle stacks, it will not work right.
   * Instead, after correcting all of the CTF parameters and 'set'ing them, exit ctfit, and use 'applyctf' to flip the phases and store the parameters:
     * applyctf will get the CTF parameters from the memories you set in ctfit (ctfparm.txt file)
     * applyctf image1.hed image1.fix.hed setparm flipphase
     * applyctf image2.hed image2.fix.hed setparm flipphase
     * ...
  * Run 'ctfit *.tnf' This will open the results of the automatic fitting.
  * If you need to correct any of the fit values (you very well may):
   * select the image in the upper right list in the control panel
* make necessary adjustments
   * press the 'set' button below the lower right list in the control panel, this will save your parameters to disk
  * Do NOT use the 'phase flip' menu item when working this way, since you loaded '.tnf' files instead of the original particle stacks, it will not work right.
  * Instead, after correcting all of the CTF parameters and 'set'ing them, exit ctfit, and use 'applyctf' to flip the phases and store the parameters:
   * applyctf will get the CTF parameters from the memories you set in ctfit (ctfparm.txt file)
   * applyctf image1.hed image1.fix.hed setparm flipphase
   * applyctf image2.hed image2.fix.hed setparm flipphase
   * ...

Q: How do I go about doing CTF correction in EMAN1

A: Much of this is described in the 4 step tutorial interface, but the description here is more up to date.

CTF correction in EMAN1, is, unfortunately, a somewhat time-consuming process. This has been massively simplified in EMAN2, but as EMAN2 still remains 'experimental' you may wish to stick with EMAN1 for a while longer for serious refinement work. It is a VERY good idea to make sure you are using a current snapshot version of EMAN1, rather than 1.8. There are 3 steps for optimally performing CTF correction in EMAN1:

  1. Determine a [:EMAN1/FAQ/StructureFactor:1-D Structure Factor] from 3-5 of your images

    • Use images over a reasonable range of defocuses
    • Don't forget to pick a good amplitude contrast (10-20% works best for cryo data)
  2. Use fitctf to fit all of your particle stacks automatically
    • use the same %AC you used above
    • guess at a B-factor based on the fitting above. For a high-end 300 keV instrument this may be as low as 50-100. For a typical 200 FEG scope it will be in the 150-200 range,

and for older/LAB6 scopes it may be as high as ~1000. Note that these are 2 different conventions for B-factor, EMAN1 B-factors are multiplied by 4 to compare with the MRC convention.

  • run this program once for each stack of particles (one stack per frame). The particles should be unmasked and unfiltered !
  • fitctf <particles from one frame> sf=<sf file from above> apix=<A/pix> V=<voltage> Cs=<spher aber> acfix=<amp contrast, 0.1 is 10%> Bhint=<guess at EMAN1 style B-factor>

  1. Go back and [:FAQ_EMAN_USING_23:manually] check your results
    • Run 'ctfit *.tnf' This will open the results of the automatic fitting.
    • If you need to correct any of the fit values (you very well may):
      • select the image in the upper right list in the control panel
      • make necessary adjustments
      • press the 'set' button below the lower right list in the control panel, this will save your parameters to disk
    • Do NOT use the 'phase flip' menu item when working this way, since you loaded '.tnf' files instead of the original particle stacks, it will not work right.
    • Instead, after correcting all of the CTF parameters and 'set'ing them, exit ctfit, and use 'applyctf' to flip the phases and store the parameters:
      • applyctf will get the CTF parameters from the memories you set in ctfit (ctfparm.txt file)
      • applyctf image1.hed image1.fix.hed setparm flipphase
      • applyctf image2.hed image2.fix.hed setparm flipphase
      • ...

Now you can continue with processing. Typically:

mkdir refine cd refine lstcat.py start.hed ../*.fix.hed lstfast.py start.hed cp start.hed start.img (make threed.0a.mrc) refine ...

Note that the above commands create a start.hed/img file which is not really in imagic format. Rather it's a text file linking to the *fix.hed particles you created, so it's very important that you not rename or remove those files. EMAN1 will treat these text files as if they were real image files.